10.2 Angle measurement
10.2 Angle measurement
If you want to measure angles send the following parameters:
Product Parameter (CMD_RECEIVE_PRODUCT_DATA - ID 40)
J00_ProductType = 1 J00_PolygonNr = 2 J00_PolygonPoint_X00_MM = -500.0 J00_PolygonPoint_Z00_MM = 0.0 J00_PolygonPoint_X01_MM = 500.0 J00_PolygonPoint_Z01_MM = 500.0
Sensor Parameters (CMD_RECEIVE_SENSOR_DATA - ID 43)
DataMode = 8 NbrLines = -1
You will permanent get a result strings like:
JOB 0.0001: Phi=156.36 Cx=00.4011 Cy=-0.9161 B=-0104.19 LAlgo=2 Points=0501
DetLen=042.6 Filter=100 dPhi=0.03 Error=0 Sh=0012us Gain=000
TimeStamp=0000006468167ms
For accuracy reason ignore the first 5-10 angles after a new plate or ROI is defined, as the auto exposure takes some images for adaption.